Protecting Process Analyzers with Parker Balston Sample Filters

This article covers continuous on-line process analyzer sampling: conditioning a plant gas or liquid stream before measurement. It does not prescribe gas-supply or sample-preparation filters for benchtop laboratory instruments. Parker Balston offers process-sample filter media and housings for different separations; the correct choice depends on sample chemistry, phase, flow, pressure, temperature and allowable delay.

Key Takeaways

  • Parker’s process-analyzer guidance identifies contamination control, representative samples, serviceability and low sample delay as key design considerations for continuous plant sampling.
  • For process sampling, filter media and wetted materials must be compatible with the sample and must not remove or alter the component being measured.
  • Parker rates Balston Microfibre grades from 93% (DX) to 99.99999+% (AAQ) retention at 0.01 micron in gas.
  • For coalescing, Parker advises choosing the grade for drainage rate, not the highest efficiency.
  • A slipstream (bypass) arrangement keeps lag time low and extends element life.

This article covers why plant samples are hard on analyzers and what good sample filters have to do. It then explains how to choose element grades, housings and filter placement for gas and liquid samples.

Why plant samples are hard on analyzers

Parker’s Filtering Samples to On-Line Process Analyzers concerns continuous plant sample streams. Its comparisons with laboratory operation are context for process sampling, not performance evidence for a particular laboratory instrument. The guidance below applies to the process-sample setting throughout.

  • Process-sample systems must be designed for the actual continuous flow and contaminant loading of the plant stream.
  • Set maintenance intervals from actual service conditions and the selected filter instructions; illustrative test frequencies are not a specification.

What analyzer sample filters must do

The paper sets out the characteristics a sample filter needs beyond simply catching dirt:

  • It must not change the sample. Other than removing unwanted impurities, the filter should not alter composition, which limits media to chemically inert materials such as glass, stainless steel and PTFE.
  • It must run reliably between scheduled checks. Sample filters often sit in remote locations, so plugging or rupture between maintenance visits is a real cost.
  • It must be easy to service. Elements should be rugged, housings should have few loose parts, and it should be difficult to install an element incorrectly.
  • It must add minimal lag time. Lag time is the delay between a change in the process and the analyzer detecting it. Large dead volume in a housing slows the analyzer’s response to process changes.

For process samples, Parker describes its Balston sample-filter media as resin-bonded borosilicate microfibres, with other media and housing choices for different services. Compatibility is model- and sample-specific. The presence of glass, stainless steel or PTFE is not a universal guarantee that a measured analyte will remain unchanged.

Choosing the separation and element grade

Removing liquid from gas samples with a coalescing filter

A coalescing filter is a filter that merges fine liquid droplets into larger drops that drain away. Glass microfibre elements coalesce fine droplets into larger drops, which drain by gravity from the outside of the element to the housing drain. Because liquid drains as fast as it arrives, the element keeps its initial retention efficiency while wet. According to Parker, you should select the grade for liquid drainage rate rather than for the highest efficiency rating. Drainage slows as efficiency rises. If liquid arrives in slugs, choose a housing with a large bowl and an automatic drain.

Separating entrained liquid from process gas samples

For process gas samples carrying entrained liquid, Parker’s sample-filter guidance describes a hydrophobic membrane with a bypass flow. Gas and vapour can pass through the membrane while rejected liquid leaves via the bypass, subject to the selected membrane’s conditions. This is liquid separation, not removal of water vapour or a guarantee of a zero-water sample. A coalescing prefilter may protect the membrane where liquid or particulate loading is high.

For gas filtration in a process-sample system, Parker’s sample-filter selection table gives these grade-specific retention efficiencies at 0.01 micron. They are not liquid-filtration micron ratings or guarantees of sample representativeness:

Balston Microfibre gradeRetention efficiency at 0.01 micron (gas)
DX, DQ, DH, DS93%
BX, BQ, BH, BS99.99%
AQ99.9999+%
AAQ99.99999+%

Where to put sample filters to cut lag time

Samples are usually carried to the analyzer at a relatively high flow to keep lag time short, while the analyzer uses only a small part of that flow. Parker describes a slipstream, or bypass, arrangement. The high flow sweeps the element and the filter reservoir, and a third port draws filtered sample to the analyzer. This keeps lag time low and extends element life, because only the analyzer’s small flow is filtered.

Some common situations from Parker’s paper:

  • Stack gas: install a filter rated for the gas temperature at the start of the sample line. Add a coalescing filter after the sample is cooled. Glass bowls allow a visual check of liquid level.
  • High-pressure steam and condensate: cool the sample, and filter upstream of pressure-reducing valves to prevent particles from pitting the valve seats.
  • Coalescing near the analyzer: place the liquid coalescing filter as close to the analyzer as practical to reduce the chance of condensation between filter and instrument.

Matching analyzer sample filters to the service

Parker’s sample-filter overview groups housings by construction and sampling duty. Use the exact model’s combined pressure/temperature rating and compatible seals; separate family maximums are not an operating envelope for every model.

Process sampling needCheck on the selected filter
Low sample delayInternal volume, sample flow and slipstream arrangement
Liquid carry-over in a gasCoalescing drainage, membrane type, bypass and liquid load
High pressure or temperatureCombined rating of the complete model, bowl, seals and connections
Corrosive or trace-component sampleWetted-material compatibility and analyte recovery in the actual sample

Material options across the range include 316 stainless steel, aluminum, PTFE, Monel, carbon steel and Pyrex glass, depending on the family. Check chemical compatibility with the full sample composition, including trace components.

For Canadian process-sampling applications, see Davis Controls’ gas and liquid filtration overview and Parker Balston filtration guide. Ask Davis Controls to confirm the model, service suitability and availability.

Frequently Asked Questions

Should I always choose the highest-efficiency element?

No. For coalescing liquid from a gas, Parker advises choosing the grade for drainage rate, since higher-efficiency grades drain more slowly and can flood. Match the grade to the job.

Why does my membrane filter need a bypass flow?

In the process-gas arrangement described by Parker, rejected liquid leaves through the bypass. Follow the selected model’s flow and differential-pressure limits. Water vapour can pass through a hydrophobic membrane, so bypass separation must not be described as drying the sample gas.

How do I reduce sample lag time?

Use a slipstream arrangement, keep filter dead volume small on the analyzer line, and locate final filters close to the analyzer.